ACCURACY DIAGNOSTIC OF DIGITAL MULTIMETERS’ INDICATIONS
Roman Tabisz
rtabisz@prz.edu.plPolitechnika Rzeszowska, Zakład Metrologii i Systemów Diagnostycznych, Laboratorium Badań i Kalibracji „LABBiKAL” (Poland)
Abstract
Methods of calculating maximum permissible error (MPE) and its two components: multiplicative (PEmul) and additive (PEadd) are described. Principles of accuracy diagnostic of multimeters' indications are explained. Use of calibration results to determine corrective function that allows for reduction of systematic errors is proposed. Presented results of experimental tests prove that calculating uncertainty of measurement results obtained after the correction of systematic error - with use of Type B method - can be based solely on additive component (PEadd).
Keywords:
maximum permissible error, calibration, accuracy diagnostic, systematic errors correctionReferences
Tabisz R.A.: Walidacja przemysłowych procesów pomiarowych. Monografia. Oficyna Wydawnicza Politechniki Rzeszowskiej, Rzeszów, 2013.
Google Scholar
JCGM 200:2008 International vocabulary of metrology.
Google Scholar
JCGM 100:2008 Evaluation of Measurement data-Guide to the expression of uncertainty in measurement.
Google Scholar