Simulation of the operation of a four-channel ultrasonic flowmeter with identical signal trajectories in acoustic channels

Main Article Content

Yosyp Bilynsky

yosyp.bilynsky@gmail.com

https://orcid.org/0000-0002-9659-7221
Andrii Stetsenko

andrii.energo@gmail.com

https://orcid.org/0009-0004-2967-8867

Abstract

In this work, a four-channel ultrasonic flowmeter with identical signal paths is proposed. Various designs of ultrasonic flowmeters have been analysed, and the rationale for developing such a measuring transducer is presented. The operation of the four-channel flowmeter was simulated both in the SOLIDWORKS Flow Simulation environment and using a classical analytical approach. The simulation results are presented in terms of the measurement error at specific flow rates, considering both the presence of a flow conditioning device and standard flow resistances. A comparative analysis of the obtained results demonstrates the advantages and potential of the proposed ultrasonic flowmeter with identical signal paths.

Keywords:

four-channel ultrasonic flowmeter with identical signal paths, simulation, measurement error, flow preparation device, additional resistances

Sustainable Development Goal (SDG)

  • Industry, Innovation, Technology and Infrastructure

References

Article Details