FUNCTIONALLY INTEGRATED DEVICE FOR TEMPERATURE MEASUREMENT

Les Hotra


Lviv Polytechnic National University, Department of Electronic Engineering (Ukraine)

Oksana Boyko

oxana_bojko@ukr.net
Danylo Halytsky Lviv National Medical University (Ukraine)
https://orcid.org/0000-0002-8810-8969

Igor Helzhynskyy


Lviv Polytechnic National University, Department of Electronic Engineering (Ukraine)

Hryhorii Barylo


Lviv Polytechnic National University, Department of Electronic Engineering (Ukraine) (Ukraine)

Marharyta Rozhdestvenska


Yuriy Fedkovych Chernivtsi National University, Department of Radio Engineering and Information Security (Ukraine)
https://orcid.org/0000-0002-0333-2604

Halyna Lastivka


Yuriy Fedkovych Chernivtsi National University, Department of Radio Engineering and Information Security (Ukraine)

Abstract

The article presents a method of implementing a functionally integrated device for temperature measurement, which allows for controlled heating of the primary temperature transducer, measurement of the heating temperature as well as the temperature and differential temperature of the investigated and reference samples. The heating speed is regulated by the selection of the frequency and duration of the control impulses. To measure the temperature and temperature difference, it is proposed to use measuring currents of different polarity, which make it possible to simplify the device design. The methods of linearisation of the conversion function of primary temperature transducer based on the formation of compensating currents in given measurement ranges have been investigated. The conducted studies showed that the temperature measurement error does not exceed 0.11°C and 0.005°C in the control heating mode and in the temperature measurement mode, respectively. The temperature measurement error of the investigated and reference samples and the differential temperature measurement error does not exceed ±0.003°C and 0.001°C, respectively.


Keywords:

temperature measurement, transistor, linearisation

[1] Azadi Kenari S. et al.: Thermal Flow Meter with Integrated Thermal Conductivity Sensor. Micromachines 14(7), 2023, 1280.
  Google Scholar

[2] Babak V., Kovtun S., Dekusha O.: Information-measuring technologies in the metrological support of heat flux measurements. CEUR Workshop Proceedings 2608, 2020, 379–393.
  Google Scholar

[3] Barylo G. I. et al.: Signal transducer of functionally integrated thermomagnetic sensors. Visnyk NTUU KPI 76, 2019, 63–71.
  Google Scholar

[4] Bianchi L. et al.: Thermophysical and mechanical properties of biological tissues as a function of temperature: A systematic literature review. International Journal of Hyperthermia 39(1), 2022, 297–340.
  Google Scholar

[5] Boyko O. et al.: Functionally integrated sensors of thermal quantities based on optocoupler. Proceeding of SPIE 10808, 2018, 306–311.
  Google Scholar

[6] Boyko O., Holyaka R., Hotra Z.: Functionally integrated sensors on magnetic and thermal methods combination basis. 14th International Conference on Advanced Trends in Radioelecrtronics, Telecommunications and Computer Engineering (TCSET). 2018, 697–701.
  Google Scholar

[7] Boyko O., Hotra O.: Improvement of dynamic characteristics of thermoresistive transducers with controlled heating. Przegląd Elektrotechniczny 5, 2019, 110–113
  Google Scholar

[8] Boyko O. V., Hotra Z. Y.: Analysis and research of methods of linearization of the transfer function of precision semiconductor temperature sensors. Physics and Chemistry of Solid State 21(4), 2020, 737–742.
  Google Scholar

[9] Braissant O. et al.: Biomedical use of isothermal microcalorimeters. Sensors 10(10), 2010, 9369–9383.
  Google Scholar

[10] Braissant O. et al.: Use of Isothermal Microcalorimetry to Monitor Microbial Activities. FEMS Microbiol. Lett 303, 2010, 1–8.
  Google Scholar

[11] Conway A. et al.: Accuracy and precision of zero-heat-flux temperature measurements with the 3M™ Bair Hugger™ Temperature Monitoring System: a systematic review and meta-analysis. Journal of Clinical Monitoring and Computing 35, 2021, 39–49.
  Google Scholar

[12] Dekusha O. et al.: Information-Measuring Technologies in the Metrological Support of Thermal Conductivity Determination by Heat Flow Meter Apparatus. Systems, Decision and Control in Energy I Springer, Cham. 2020, 217–230.
  Google Scholar

[13] Farkas G.: Temperature-Dependent Electrical Characteristics of Semiconductor Devices. Theory and Practice of Thermal Transient Testing of Electronic Components 2023, 139–169.
  Google Scholar

[14] Feng J. et al.: Droplet-based differential microcalorimeter for real-time energy balance monitoring. Sensors and Actuators B: Chemical, 312, 2020, 127967.
  Google Scholar

[15] Gros S. J. et al.: Personalized treatment response assessment for rare childhood tumors using microcalorimetry–exemplified by use of carbonic anhydrase IX and aquaporin 1 inhibitors. International journal of molecular sciences 20(20), 2019, 4984.
  Google Scholar

[16] Hotra O., Boyko O., Zyska T.: Improvement of the operation rate of medical temperature measuring devices. Proc. SPIE 92914, 2014, 92910A.
  Google Scholar

[17] Hotra O., Boyko O.: Analogue linearization of transfer function of resistive temperature transducers. Proc. SPIE 9662, 2015, 966247.
  Google Scholar

[18] Hotra O., Boyko O.: Compensation bridge circuit with temperature-dependent voltage divider. Przegląd Elektrotechniczny 4a, 2012, 169–171.
  Google Scholar

[19] Hotra O., Dekusha O., Kovtun S.: Analysis of the Characteristics of Bimetallic and Semiconductor Heat Flux Sensors for In-Situ Measurements of Envelope Element Thermal Resistance. Measurement 2021, 109713.
  Google Scholar

[20] Khaw M. K., Mohd-Yasin F., Nguyen N. T.: Microcalorimeter: Design considerations, materials and examples. Microelectronic Engineering, 158, 2016, 107–117.
  Google Scholar

[21] Kuril A. K.: Differential scanning calorimetry: a powerful and versatile tool for analyzing proteins and peptides. J Pharm Res Int 36(7), 2024, 179–187.
  Google Scholar

[22] Kuttner H. et al.: Microminiaturized thermistor arrays for temperature gradient, flow and perfusion measurements. Sensors and Actuators A: Physical 27(1-3), 1991, 641–645.
  Google Scholar

[23] Lubbers B. et al.: Microfabricated calorimeters for thermometric enzyme linked immunosorbent assay in one-Nanoliter droplets. Biomedical Microdevices 21, 2019, 1–7.
  Google Scholar

[24] Lundén O. P., Paldanius T.: Linearization of BJTs with logarithmic predistortion. IEEE Radio and Wireless Symposium (RWS) 2019, 1–3.
  Google Scholar

[25] Magoń A., Pyda M.: Apparent heat capacity measurements and thermodynamic functions of d(−)-fructose by standard and temperature-modulated calorimetry. The Journal of Chemical Thermodynamics 56, 2013, 67–82.
  Google Scholar

[26] Martins E. et al.: Skin byproducts of reinhardtius hippoglossoides (Greenland Halibut) as ecosustainable source of marine collagen. Applied Sciences 12(21), 2022, 11282.
  Google Scholar

[27] Ni S. et al.: A SiN microcalorimeter and a non-contact precision method of temperature calibration J. Microel. Syst. 29(5), 2020, 1103–1105.
  Google Scholar

[28] Pertijs M. A., Makinwa K. A., Huijsing J. H.: A CMOS smart temperature sensor with a 3σ inaccuracy of ±0.1C from -55C to 125C. IEEE Journal of Solid-State Circuits 40(12), 2005.
  Google Scholar

[29] Pollock D. D.: Thermocouples: theory and properties. Routledge 2018.
  Google Scholar

[30] Sundararajan S. et al.: BJT Based Translinear Implementation of an Evolutionary Optimised Non-Linear Function for Sensor Linearisation. IETE Journal of Research 70(6), 2023, 5905–5918.
  Google Scholar

[31] Wang F., Han Y., Gu N.: Cell temperature measurement for biometabolism monitoring ACS Sens. 6(2), 2020, 290–302.
  Google Scholar

[32] Wang Y. et al.: Recent advances of microcalorimetry for studying cellular metabolic heat. TrAC Trends in Analytical Chemistry 143, 2021, 116353.
  Google Scholar

[33] Zaporozhets A. et al.: Information Measurement System for Thermal Conductivity Studying. Advanced Energy Technologies and Systems I. Studies in Systems, Decision and Control 395, 2022.
  Google Scholar

[34] Zhu H. et al.: The development of ultrasensitive microcalorimeters for bioanalysis and energy balance monitoring. Fundamental Research 4(6), 2023, 1625–1638.
  Google Scholar

Download


Published
2024-12-21

Cited by

Hotra, L., Boyko, O., Helzhynskyy, I., Barylo, H., Rozhdestvenska, M., & Lastivka, H. (2024). FUNCTIONALLY INTEGRATED DEVICE FOR TEMPERATURE MEASUREMENT. Informatyka, Automatyka, Pomiary W Gospodarce I Ochronie Środowiska, 14(4), 32–37. https://doi.org/10.35784/iapgos.6720

Authors

Les Hotra 

Lviv Polytechnic National University, Department of Electronic Engineering Ukraine

Authors

Oksana Boyko 
oxana_bojko@ukr.net
Danylo Halytsky Lviv National Medical University Ukraine
https://orcid.org/0000-0002-8810-8969

Authors

Igor Helzhynskyy 

Lviv Polytechnic National University, Department of Electronic Engineering Ukraine

Authors

Hryhorii Barylo 

Lviv Polytechnic National University, Department of Electronic Engineering (Ukraine) Ukraine

Authors

Marharyta Rozhdestvenska 

Yuriy Fedkovych Chernivtsi National University, Department of Radio Engineering and Information Security Ukraine
https://orcid.org/0000-0002-0333-2604

Authors

Halyna Lastivka 

Yuriy Fedkovych Chernivtsi National University, Department of Radio Engineering and Information Security Ukraine

Statistics

Abstract views: 11
PDF downloads: 18


License

Creative Commons License

This work is licensed under a Creative Commons Attribution 4.0 International License.