APPLICATION OF DIGITAL IMAGE PROCESSING METHODS FOR OBTAINING CONTOURS OF OBJECTS ON ULTRASOUND IMAGES OF THE HIP JOINT
Pavlo Ratushnyi
ratushnyj.p.m@vntu.edu.uaVinnytsia National Technical University, Department of Electronics, Vinnytsia, Ukraine (Ukraine)
https://orcid.org/0000-0002-3365-8699
Yosyp Bilynskyi
Vinnytsia National Technical University, Department of Electronics, Vinnytsia, Ukraine (Ukraine)
https://orcid.org/0000-0002-9659-7221
Stepan Zhyvotivskyi
Vinnytsia National Technical University, Department of Electronics, Vinnytsia, Ukraine (Ukraine)
http://orcid.org/0000-0002-2379-0645
Abstract
In this work, the problems of research of ultrasonic images of joints are formulated. It is that for early diagnosis of developmental disorders of the hip joints needs to take frequent pictures, and the least harmful to health is ultrasound. But the quality of such images is not sufficient for high-quality automated measurement of geometric parameters and diagnosis of deviations. The ultrasound image of the hip joint is evaluated by quantifying the exact values of the acetabular angle, the angle of inclination of the cartilaginous lip, and the location of the center of the femoral head. To get these geometric parameters, you need to have clear images of objects. And for the operation of automated computer measurement systems, it is necessary to use such methods of pre-digital image processing, which would give clear contours of objects. Known and available image processing algorithms, in particular contour selection, face problems in processing specific medical images. It is proposed to use the developed method of sharpening to further obtain high-quality contour lines of objects. A mathematical model of the method is presented, which is a formula for converting the intensity values of each pixel of a digital image. As a result of this method, the noise component of the image is reduced, and the intensity differences between the background and the objects are increased, and the width of these differences is one pixel. The algorithm of a sequence of processing of ultrasonic images and features of its application have resulted. The results of the developed set of methods are given. The paper presents the results of processing the real image of the hip joint, which visually confirms the quality of the selection of objects on view.
Keywords:
ultrasound images, contour, sharpening, digital processing, hip joint, diagnosticsReferences
Bilynsky Y., Horodetska O., Ratushny P.: Prospects for the use of new methods of digital processing of medical images. 13th International Conference on Modern Problems of Radio Engineering, Telecommunications and Computer Science TCSET, 2016, 780–783 [http://doi.org/10.1109/TCSET.2016.7452182].
DOI: https://doi.org/10.1109/TCSET.2016.7452182
Google Scholar
Bilynsky Y., Ratushny P., Klimenko І.: The method of sharpening low-contrast two-dimensional images. Bulletin of the Polytechnic Institute 6/2009, 12–15.
Google Scholar
Bilynsky Y., Ratushny P., Melnichuk A.: The method for improving image sharpness. Applicant and patentee Vinnytsa National Technical University 200907326, Pat. 45887, Ukraine, G 06 K 9/36. applic. 13.07.09; publ. 25.11.09, bull. 22.
Google Scholar
Bilynsky Y., Yukysh S., Ratushny P.: Edge detection detector based on low-pass filtering. Bulletin of Khmelnytsky National University 1/2009, 230–223.
Google Scholar
Bilynsky Y.Y. et al.: Contouring of microcapillary images based on sharpening to one pixel of boundary curves. Proc. SPIE 10445, 2017, 104450Y [http://doi.org/10.1117/12.2281005].
DOI: https://doi.org/10.1117/12.2281005
Google Scholar
Bilynsky Y.Y. et al.: Controlling geometric dimensions of small-size complex-shaped objects. Proc. SPIE 10445, 2017, 104450I [http://doi.org/10.1117/12.2280899].
DOI: https://doi.org/10.1117/12.2280899
Google Scholar
Nikolskyy A. I. et al.: Using LabView for real-time monitoring and tracking of multiple biological objects. Proc. SPIE 10170, 2017, 101703H [http://doi.org/10.1117/12.2261424].
DOI: https://doi.org/10.1117/12.2261424
Google Scholar
Ratushny P. M. et al.: Research of the mask size for the method of increasing the sharpness to the maximum slope of the boundary curve. Measuring and computing equipment in technological processes. Khmelnytskyi 3/2014, 71–74.
Google Scholar
Authors
Pavlo Ratushnyiratushnyj.p.m@vntu.edu.ua
Vinnytsia National Technical University, Department of Electronics, Vinnytsia, Ukraine Ukraine
https://orcid.org/0000-0002-3365-8699
Associate Professor of Electronics and Nanosystems at Vinnytsia National Technical University. In 2011 defended his dissertation on "Methods and system of low-contrast image processing for the evaluation
of microcapillaries of human limbs" in the specialty "Biological and medical devices and systems".
The main scientific direction is computer processing of biological and medical images for research
of geometrical parameters of objects.
Authors
Yosyp BilynskyiVinnytsia National Technical University, Department of Electronics, Vinnytsia, Ukraine Ukraine
https://orcid.org/0000-0002-9659-7221
Associate Professor of Electronics and Nanosystems at Vinnytsia National Technical University. He defended his doctoral dissertation in 2009. He has more than 250 scientific works, including 65 in professional publications, 65 patents, 4 monographs, 10 educational and methodological.
Authors
Stepan ZhyvotivskyiVinnytsia National Technical University, Department of Electronics, Vinnytsia, Ukraine Ukraine
http://orcid.org/0000-0002-2379-0645
Statistics
Abstract views: 277PDF downloads: 220
License
This work is licensed under a Creative Commons Attribution-ShareAlike 4.0 International License.